INSTALLING THE CANTILEVER
31
- Remove the DropStop.
The laser beam is now unblocked, and the Probe Status light on the easy-
Scan 2 Controller should now stop blinking red. If this is not the case refer
to Chapter Problems and Solutions (p.66).
Mounting the cantilever: top left: Inserting the cantilever insertion tool; top right: Inserting/
Removing the cantilever; bottom right: Correctly inserted cantilever
Cantilever Alignment: left: Correct, the mirrored environment light shows a pattern that is
continuous on the cantilever and the alignment chip; centre, right: Incorrect, the mirrored
environment light shows a different pattern on the cantilever than on the alignment chip.